SENSOR
Technology for Human life
SENSOR
Technology for Human life
Electrochemical
0 - 2,000 ppm
Applications
• Ideal for portable instrument
• Industry safety (ambient air monitoring appllications)
• Residential CO alarms
• Fire alarms
• Etc.
|
Performance Characteristics |
|
|---|---|
| Output Signal | 70±20 nA/ppm |
|
Typical Baseline Range (pure air) |
≤±2 ppm CO equivalent |
|
Filter Capacity |
≥20,000 ppm hours |
| T90 Response Time |
<20 seconds |
|
Measurement Range |
0-2,000 ppm |
|
Maximum Overload |
5,000 ppm |
|
Linearity |
Linear up to 2,000 ppm |
|
Repeatability |
≤±2 ppm CO equivalent |
|
Recommended Load Resistor |
10 ohms |
|
Resolution (Electronics Dependent) |
≤0.5 ppm typical |
| Environmental | |
|
Temperature Range Continuous |
-20℃ to +50℃ |
|
Pressure Range |
800 to 1200 mbar |
|
Operating Humidity Range |
15% to 90% Rh |
|
Life Time |
|
| Long Term Output Drift |
≤5% per annum |
|
Recommended Storage Temp. |
0℃ to 20℃ |
|
Expected Operating Life |
≥24 months in air |
|
Standard Warranty |
24 months from the date of dispatch |
|
Intrinsic Safety Data |
|
| Maximum at 10,000ppm |
80mA |
|
Maximum o/c Voltage |
<1.0V |
|
Maximum s/c Current |
<0.1A |